사 양 |
Inspection Technology |
|
Eddy Current. |
Operating Frequencies |
|
60 kHz, 120 kHz, 240 kHz, 480 kHz, 960 kHz. |
Conductivity Range |
|
0.5% IACS to 110% IACS, 0.28-64 MS/m |
Accuracy |
|
At 20℃. At 10% IACS: ±0.1% IACS.
At 100% IACS: ±0.5% IAC
Over Range 0-40℃: At 10% IACS:
±0.2% IACS. At 100% IACS: ±0.8% IACS
Probe in thermal equilibrium with metal. |
Display Resolution |
|
Up to 3 decimal places |
Lift Off |
|
13mm probe compensated to 0.020"
(0.5mm) 7mm probe compensated to 0.010" (0.25mm) |
Temperature Measurement |
|
In-probe sensor (accurate to 0.5℃)
Range 0℃ to + 50℃ |
Automatic Temperature Compensation |
|
Conductivity measurements are corrected to the 20℃ value. |
Environmental Range |
|
0 to 95℃ relative humidity, 0℃ to + 50℃ for reliable operation |
Display |
|
2.8" (70mm) 320 x 240 pixels colour display-LCD with selectable backlight. |
Construction & Storage |
|
High impact, splash-proof, moulded UL94-5VA flame-retardant ABS case.
Protective rubber boot to protect the unit, probes, probe cabel,
operator manual on USB, and removable stand. |
Conductivity Standards |
|
On top of unit.
Removable for value verification, and when attached ensures
thermal equilibrium. |
Power |
|
2 x 1.5 V AA NiMH Batteries, Approx up to 6 hrs life.
Can also use non-rechargable AA cells. |
Size |
|
163mm Long x 80mm Wide x 25mm Deep |
Weight |
|
350g (0.77 pounds) including batteries |
Data Logger Memory |
|
Removable 2GB micro SD Card allowing over 1 million readings
to be stored. |
PC Connectivity |
|
USB port for charger and PC communications |
Probes |
|
12.7mm diameter for 60 kHz to 480 kHz.
7mm probe operates at 480 kHz & 960 kHz.
Probes are interchangeable with simple operator resetting procedure.
Probes are field exchangable and do not require return to manufacturer
for calibration. |
Accessories |
|
Settings Reference Blocks - A range of
conductivity references standards tracable to US
and European standards are available for in-field use.
Up to five can be mounted on an aluminium anodised holding plate. |